MULTI ANALYZER DETECTOR UNIT |
|
Measuring system |
Infrared
absorption type |
|
Measuring wavelength |
Up to 10
wavelengths |
|
Measuring scope |
Up to 4
constituents |
|
Source of light |
Tungsten
lamp |
|
Measuring distance and measuring diameter |
30/200mm
or 50/300mm |
|
Output signal |
|
1) |
Analog signal: 4 to 20mADC; ±0.2% of full scale
(Load resistance: less than 500) |
|
2) |
Communications signal: RS-485 MODBUS (STD) |
|
3) |
Ethernet (LAN) |
|
|
Output renewal cycle |
28ms |
|
Display & setting |
LED
display shows the
measurement data and
setting value.
Various parameters are settable by keypad. |
|
Computing function |
Ratio
calculation for 2-color or 3-color, and multiple regression calculation. |
|
Calibration curve number |
up to 99
curves |
|
Calibration curve |
Linear, quadratic, cubic and
multiple regression line
Calibration curve correction:
Linear & quadratic |
|
Smoothing |
0 to 99
seconds, optional setting |
|
Calibration |
By
checking plate |
|
Setting detector number |
Use
multi-head then set the
detector number by keypad |
|
Setting channel number |
Set
calibration curve number
by keypad |
|
Self-diagnostic function |
Display
and output alarm by digital communications and/or contact signal |
|
Input correction |
Corrects
the measured data by
external 4 to 20mA DC (1 input)
(sample temperature correction, etc.) |
|
External Di/o |
Di(contact input), selectable from preset,
data-hold, real/ smooth functions.
Do(contact output), selectable from
self-diagnostic function (1b) or upper/ lower
limit alarm (1a). |
|
Ambient temperature range |
0 to 50 °C (air cooling is necessary for higher than
45°C) |
|
Power supply |
24V DC |
|
Power consumption |
Approx. 30VA |
|
Connection |
Terminals connection |
|
Casing |
Aluminum casing,
drip-proof structure (conforming to IEC529, IP65) |
|
Weight |
Approx. 4.3kg |
|
Mounting |
Suspension the analyzer uses 4-M8 bolts. |
|
CE-marking specifications |
EN61326+A |
|
SETTING DISPLAY UNIT |
|
Detector unit input |
RS-485, Connectable with max.9 detector units |
|
Analog output |
|
1) |
Analog signal: 4 to 20mADC; 2 output*
(Load resistance: less than 500)
* In the case of multi-head, output from No.1 and
No.2
connected detectors. |
|
2) |
Communications signal: specified from RS-232C,
RS-422A or RS-485. |
|
|
Output scaling |
By
numeric key |
|
Output renewal cycle |
Communications output: 28ms ´ detector number |
|
Display |
|
1) |
Measured data, LED 5-digit |
|
2) |
Head No., CH. No., parameter. |
|
|
Setting detector number |
Set
detector head No.1 to No.9 by key |
|
Setting channel number |
Set
calibration curve number by key |
|
Smoothing time |
T=0.1 to
99.9 seconds |
|
Calibration |
After
output checking plate inserted, calibrate by key or external contact. |
|
Hold/preset |
Hold or
preset the display and output by
key or external contact. |
|
Calibration curve correction |
Corrects
the calibration
curve online or linear,
quadratic correction. |
|
External setting |
Head
number, CH. No., Calibration Hold,
Preset. |
|
Alarm function |
Contact 1
output (HCL) outside of setting-range |
|
Self-diagnostic |
Contact 1 output |
|
Power supply |
100 to 240V AC 50/60Hz |
|
Power consumption |
max. 15VA |
|
Working ambient temperature |
0 to 50°C |
|
Casing |
ABS resin (IP65) |
|
Mounting |
Panel-mount type |
|
Weight |
Approx. 0.6kg |
|
FIBER UNIT |
|
Measuring distance and measuring diameter |
|
with lens: |
f25/25mm to f40/100mm |
|
without lens: |
f20/15mm to f50/50mm |
|
|
Fiber length |
Standard
1.5m, Max.5m (Reflection type fiber)
Standard 2m, Max.10m (Transmission type fiber) |
|
Fiber protection |
Stainless
steel snake tube |
|
Minimum bending radius |
R100mm |
|
Working temperature range |
0 to
150°C |
|
Purge air flow |
|
with lens: |
non air purge |
|
without lens: |
5 to 20NL/min |
|
|
Accessory |
Vertical
mounting holder,
Flange holder |