Measuring points |
Pen-writing type |
1-pen,
2-pen, 3-pen or 4-pen |
|
Dot-printing type |
6
points, 12 points, 24 points |
|
Input
signals |
Universal input |
|
Thermocouples |
B, R,
S, K, E, J, T, N, W-WRe26, WRe5-WRe26,
PtRh40-PtRh20, NiMo-Ni, CR-AuFe, Platinel II, U,
L |
|
Resistance thermometers |
Pt100
(IEC751-1995), Pt100 (IEC751-1983), JPt100,Pt50,
Pt-Co |
|
DC
voltage |
±13.8mV, ±27.6mV, ±69.0mV, ±200mV, ±500mV,
±2V,±5V, ±10V, ±50V, ±50V |
|
DC current |
By
shunt resistor |
|
Contact
input (Option in pen-writing type, operational
record) |
|
Accuracy rating |
±0.1%
of measuring range ± 1 digit (DC voltage,
thermocouple, resistance thermometer
inputs) |
|
Measuring cycle |
Pen-writing type |
About
100ms |
|
Dot-printing type |
About 5
seconds/6-point, about 10 seconds/12-point, about
20 seconds/24-point Option … About 1
second/6-point, about 2 seconds/12-point, about 4
seconds/24-point |
|
Reference junction compensation
accuracy |
K, E,
J, T, N, Platinel II ………………………………… Maximum
±0.5°C R, S, W-WRe26, WRe5-WRe26, NiMo-Ni,
CR-AuFe, U, L …………………………………………………………… Maximum
±1.0°C (The above errors are added to the
accuracy ratings for an internal reference
junction compensation.) |
|
Burnout
|
For
thermocouple and resistance thermometer inputs
(up-scale, down-scale or burnout
disable) |
|
Terminal board |
Detachable |
|
Printing system |
Pen-writing type |
Analog tracing |
Disposal cartridge pen |
|
Digital printing |
Plotter pen |
|
Dot-printing type |
Wire-dotting system 6-color
ribbon |
|
Printing colors |
Pen-writing type |
1st-pen
red, 2nd-pen green, 3rd-pen blue, 4th-pen
brown |
|
Dot-printing type |
1 red,
2 black, 3 blue, 4 green, 5 brown, 6
purple |
|
Chart
|
Fan-fold type, effective width 180mm, total
width 200mm, total length 20m |
|
Chart
speed |
Pen-writing type |
1 to
600mm/h, 1 to 200mm/min (default -
25mm/h) |
|
Dot-printing type |
1 to
1500mm/h (default - 25mm/h) |
|
Skip
function |
No
display or printings of channels |
|
Subtract printing |
Printing of difference of measured values
of one channel and other channel (or reference
value) |
|
Pen-lift function |
By
RECORD OFF key, all pens are lifted up
simultaneously. A lever for manually lifting
up/down of all pens is provided. |
|
Alarm
type |
Absolute value, differential,
rate-of-change (Absolute value/standby and
differential/standby are additionally available in
pen-writing types. Each 2 levels) |
|
Alarm
programming |
Individual programming each channel,
Maximum 4 levels/channel (output:
option) |
|
Rated
power voltage |
100 to
240VAC 50/60Hz, universal power
supply |
|
Power
consumption |
Pen-writing type |
About
60VA |
|
Dot-printing type |
About
45VA |
|
Environmental conditions |
Reference operation condition |
Ambient
temperature/humidity range: 21 to 25°C/45 to 65%RH
Power voltage: 100VAC Ù} 1% Attitude: Left/right 0
degree, forward tilting 0 degree, backward tilting
0 degree Warm-up time: 30minutes or
longer |
Normal operation condition |
Ambient
temperature/humidity range: 0 to 40°C/20 to 80%RH
Power voltage: 90 to 264VAC Attitude: Left/right 0
to 10 degrees, forward tilting 0 degree, backward
tilting 0 to 30 degrees |
|
Mounting |
Panel
mounting |
|
Dimensions |
Pen-writing type |
W288 x
H288 x D220mm (standard
specifications)
W288 x
H288 x D247mm (with contact mechanical relay) |
|
Dot-printing type |
W288 x
H288 x D220mm (standard
specifications)
W288 x
H288 x D247mm (with contact mechanical relay |
|
Weight
|
Pen-writing type |
About
4.0kg (4-pen/with full options) |
|
Dot-printing type |
About
3.0kg (with full options) |
|
Standards |
CE
|
EN61326
+ A1 Class A, EN61000-3-2 + A14 EN61000-3-3,
EN61010-1 + A2 |
|
UL
|
UL3111-1 (pen-writing type - approval
pending) |
|
CSA (C-UL) |
C22.2,
No. 1010 (pen-writing type - approval
pending) |
|
IP |
IEC529
IP54 (front protection) |