| Input Signals |
 |
|
|
Measuring points : |
12, 24 and 36 points |
|
Input : |
Multi-channel data range
DC voltage --- ±10mV, ±20mV, ±40mV, ±80mV, ±1.25V,
±2.5 V, ±5 V, ±10V DC current --- Shunt resistor
(100Ω, 250Ω) needs to be mounted externally
Thermocouple --- B, R, S, K, E, J, T, N,
PtRh40-PtRh20, NiMo-Ni,WRe5 – WRe26, W-WRe26,
Platinel II, U, L
Resistance thermometer --- Pt 100, JPt 100 |
|
Range setting: |
Input type and range are set with front keys |
|
Scale setting: |
The minimum and maximum values and unit are set for
each point with front keys Setting range -30000 to
30000
Decimal points Optional setting |
|
Indication accuracy: |
Refer to items of measuring ranges, accuracy rating
and display resolutions |
|
Temperature drift: |
0.1% FS/ 10°C |
|
Measuring period: |
0.1 sec/ channel |
|
Reference junction compensation accuracy: |
K, E, J, T, N, Platinel II --- ±0.5°C or less (0℃°C
or more when measuring) R, S, WRe5-WRe26, NiMo-Ni,
U, L --- ±1.0°C or less (Only when the ambient
temperature is 23°C±5°C) |
|
Input resolution: |
Approx. 1/40000 (Standard range conversion) |
|
Burnout: |
Select with/ without burnout for each input
|
|
Allowable signal source resistance: |
Thermocouple inputs, DC voltage input (10mV) ---
500Ω or less (without burnout) DC Voltage input
(except 10mV) --- 100Ω or less Resistance
thermometer inputs --- 10Ω or less/ line Three lines
are common, Pt100, JPt100 |
|
Input resistance: |
Thermocouple input, DC voltage input --- approx.1MΩ |
|
Maximum input applied voltage: |
±20V DC |
|
Input correction: |
Zero/span correction and shift correction for each
channel |
|
Maximum common mode voltage:
|
30V AC (support LVD) *250V AC at evaluation
|
|
Common mode rejection ratio: |
130dB |
|
Series mode rejection ratio: |
50dB (Only when the peak value of noise is below
standard range.) |
|
Terminal board: |
Detachable type, removable for wire connection |
| Recording
Specifications |
 |
|
|
Recording system : |
Raster scan system, 10-color wire dot printing
|
|
Recording and recording color : |
Analog recording --- color can be specified for each
channel as required. 10 colors (red, purple-red,
orange, brown, green, yellow-green, blue-green,
purple, purple-blue, black) Digital recording and
logging recording - Black Message printing --- Black
List printing --- Black |
|
Chart paper: |
Fan-fold type, Overall width 318 mm, total length
20m; Effective recording width 250mm (analog
recording) |
|
Chart speed: |
1 to 1500mm/H (in 1mm/H steps) |
|
Skip function: |
Analog recording, digital recording and digital
display can be set independently from recording
slip. |
|
Recording compensation: |
Independent setting of zero spans are available. |
| Display Specifications |
 |
|
|
Digital display: |
Color LCD panel RGB (640 x 240 dot) Display size
W149.8 x H57.4 mm |
|
Setting display: |
Common to digital display |
|
Display contents: |
Digital display
Channel display ( One-point/ multiple points
continuous/sequential indication change) Display
measuring value of each channel ( One-point/
multiple points continuous/sequential indication
change) Clock display (Hour/Minute/Second/Tag/Unit)
Chart speed display |
|
Status display: |
RECORD ON ( lights during recording) LED KEY LOCK (
lights during key lock) ALARM ( lights during alarm
activated) LED CHART END ( lights just before record
ending) FAIL ( lights during unit abnormal time) *
Sharing LED and setting display |
| Alarm Specifications |
 |
|
|
Alarm display : |
Occurrence CH No., data is displayed in red when
alarm occurs |
|
Alarm types: |
High limit, low limit |
|
Alarm setting method: |
Individual setting for each point four levels/
channels |
Alarm output:
(Option) |
See option specification |
| General Specifications |
 |
|
|
Rated power voltage: |
100 to 240V AC (universal power supply) 50/60Hz |
|
Maximum power consumption: |
100V A |
|
Reference operating condition:
|
Ambient temperature/ humidity range: 21 to 25 °C, 45
to 65%RH Power voltage: 90 to 264V Power frequency:
50/60Hz ±2% Attitude: Forward/ Backward/ Left/ Right
within 3° Warm-up time: 1 hour or longer |
|
Normal operating condition:
|
Ambient temperature/humidity range 0 to 40°C, 20 to
80% RH Power voltage: 90 to 264V Power frequency:
50/60Hz ±2% Attitude: Forward/ Backward/ Left/ Right
within 3° |
|
Transportation condition:
|
At the packed condition on shipment from our factory
Ambient temperature/ humidity range: -20 to 60 °C, 5
to 90%RH (No dew condensation) Vibration: 10 to 60
Hz, 4.9m/ S2(0.5G or less) Impact: 392m/S2 (Approx.
40G or less) |
|
Storage condition: |
Ambient temperature -20 to 60 °C, 5 to 90%RH (No dew
condensation) |
|
Working condition: |
Working temperature range 0 to 40 °C |
|
Power failure protection: |
Programmed parameters stored into EEPROM memory
Clock circuit sustained for 5 years or longer by a
lithium battery (at the operation of 8 hours or
longer per day) |
|
Insulation resistance:
|
Between primary terminals and protective conductor
terminals --- 20MΩ or more at 500V DC Between
secondary terminals and protective conductor
terminals --- 20MΩ or more at 500V DC Between
primary terminals and secondary terminals --- 20MΩ
or more at 500V DC |
|
Dielectric strength: |
Between primary terminals and protective conductor
terminals --- 1 minute at 1500V AC Between secondary
terminals and protective conductor terminals --- 1
minute at 500V AC Between primary terminals and
secondary terminals --- 1 minute at 1500V AC Note 1:
Primary terminals: power terminal, alarm output
terminal, output relay terminal Secondary terminals:
measuring input terminal, communication terminal,
external drive terminal Note 2: When testing
insulation resistance and dielectric strength,
please short-circuit every terminals of primary and
secondary terminals before the test. Test without
short-circuiting terminals can damage instruments.
|
|
Case assembly material:
|
Door (frame) --- ABS resin, Front panel --- Soda
glass, Back case --- Normal steel |
|
Color: |
Door(frame) --- White (Equivalent to DIC546 1/2),
Front panel --- Transparent, Back case --- White
(Equivalent to DIC546 1/2) |
|
Mounting: |
Panel mounting |
|
Weight: |
About 15kg (Full option) |
|
Dimensions, panel cut: |
W400 x H260 x D300 mm (Dimensions) 388 x 248mm
(Panel cut) |
|
Terminal screws: |
Measuring input, alram terminals --- M3.5 Power,
protective conductor terminal, external drive
terminal, communication terminal --- M4 |
|
Chart paper illumination : |
White LED |
| Optional
Specifications |
 |
|
|
External drive |
Chart 3- speed, chart stop, data printing, list
printing, message printing 5 types, operation
recording |
|
Alarm output |
Mechanical relay ---- 12, 24, 36 points output, max
contact capacity of 100 to 240V AC, 3A resistance
load |
|
Communication interface
|
RS-422A or RS-485 + Ethernet + 1a contact output (1a
contact output is contact output of mecha relay
|
|
Chart end output |
CHART END relay output when chart paper ended
(communication interface is required) |
|
FAIL output |
FAIL relay output when abnormality ( communication
interface is required) |
|
Receiving resistance for current input |
250Ω( for 20mA) or 100Ω(for 50mA) are externally
mounted to measure current |
|
Communication
interface specification |
 |
|
|
|
|
With communication interface
|
Without communication interface
|
|
Ethernet |
Specification |
Ethernet10BASE-T/ 100BASE-T, automated recognition,
TCP, IP, HTTP, exclusive protocol |
------------- |
|
Function |
Data display, parameter setting, with browser Data
display, parameter setting on exclusive application |
------------- |
|
RS-422ARS-485 |
Specification |
RS-422A, RS-485, Communication protocol: MODBUS
Communication specification: 9600 bps to 19200 bps
7E1 to 8N2 |
------------- |
|
Function |
Data display and parameter setting using exclusive
application |
------------- |
|
USB |
Specification |
Data display and parameter setting using exclusive
application |
|
Function |
Parameter setting for exclusive application |