LE Series Intelligent Recorders

The LE (LE1000, LE2000) series 250mm intelligent recorders accept multi-range inputs. They incorporated various superior facilities; high accuracy, 0.05%; high-speed scanning, 30 points per second; and high-speed recording, 1 line in 2 seconds. Operation is simple with front panel keys while setting-up can be executed from a personal computer or with an IC memory card.   LE Series
Features Models Specifications Accessories

Measuring points

10, 20 or 30 points
Inputs Thermocouples R, S, B, K, E, J, T, N, WWRe5-26, WWRe0-26, PR20-40, PR5-20, Ni-NiMo, AuFe-CR, Platinel, U, L
DC Voltage ±12.5mV, ±25mV, ±60mV, ±120mV, ±200mV, ±500mV, ±2V, ±5V, ±20V*, ± 50V*
DC current Applicable by adding shunt resistors (100 , 250 )
Resistance thermometers

Pt100, JPt100, Pt50, Pt-Co

  • Thermocouple and DC voltage multi-range models have a total of 37 ranges including 29 thermocouple ranges and 8 DC voltage ranges (excluding that marked with * ±20V and ±50V)
  • Thermocouple, DC voltage and resistance thermometer multi-range models have a total of 47 ranges including 29 thermocouple ranges, 8 resistance thermometer ranges and 10 DC voltage ranges.
Measuring period 30 points/second (2 seconds for maths configuration on LE2000)
Accuracy rating ±0.05% of measuring range ± 1 digit (at DC voltage ±25mV)
Reference junction compensation accuracy K, E, J, T, Nicrosil-Nisil, Platinel: ± 0.5°C
R, S, Ni-NiMo, AuFe-Cr, WWRe5-26, WWRe0-26, U, L: ± 1°C
Burnout Thermocouple input and resistance thermometer input only (DC voltage input is impossible.)
Up scale burnout
Detection pulse width of 5 ms and detection ON-OFF switching possible (internal DIP switch).
Chart Fan-fold type, total width 318mm, total length 20m, effective width 250mm (analog recording)
Printing points Analog printing Input signal: 30 points
Calculated result: 30 points
Digital printing 60 points
Logging printing 60 points
Printing system Raster scan system, 10-color wire dot printing
Printing colors Analog printing Color can be specified for each channel as required. 10 colors (red, purple-red, orange, brown, green, yellow-green, blue-green, purple, purple-blue, black)
Digital printing and logging printing Black
Chart speed 1 to 1500mm/h (in 1 mm/h steps)
Recording cycle (interval) Analog printing Chart speed interlocking cycle or arbitrary set cycle (minimum about 2 seconds)
Digital printing Arbitrary hour and minute setting (within 24 hours)
Logging printing Arbitrary hour and minute setting (1 minute to 24 hours)
Minimum interval Different depending on the number of recording channels and recording format)
Start time Arbitrary time setting (within 24 hours)
Display section Fluorescent display tube (5 x 7 dots, 17 characters, fixed status), 2 colors (blue-green and amber)
Internal display Cold-cathode discharge tube
Maths function LE1000 Subtraction, maximum, minimum, average, totalization (option)
LE2000 Special linearizer, four basic arithmetic function, square root, logarithm, minimum, maximum, average, totalizer, logical calculation, etc.
Maths input
(LE2000)
Analog input Measured data
Digital input Chart end, operational recording input
Data communications input (with optional communications interface) Analog and digital (ON/OFF)
Maths output
(LE2000)
Number of channels Up to 30 (channel Nos. 31 to 60)
Analog recording, digital recording, data display, communications output data (with optional communications interface)
Operation command Data print, message print, printing ON/OFF
Card function (using an IC memory card) Setting Parameters
Data memory Storage of measured data and calculated data
Trigger condition (1) Time, (2) Alarms, (3) End of chart, (4) Remote contacts Pretrigger value: 0 to 100% (10% step)
Alarm type Individual setting, high limit, low limit
Alarm settings Individual setting for each point (with key operation)
4 levels/channel
Alarm output 2 points by relay (AND or OR output can be specified)
Contact capacity: 100VAC, 0.5A or 200VAC, 0.2A
Power supply 81 to 264VAC (universal power supply)
Power consumption About 80VA
Operating temperature 0 to 40°C
Operating humidity 20 to 80%RH (no dew condensation)
Weight About 20kg