Type |
Wall-mount type |
Separated type |
Separated heat resistive type |
Insertion type (soft type) |
Insertion type (hard type) |
|
Model |
HN-CQ |
HN-CFA |
HN-CGA |
HN-CV |
HN-CP |
|
Humidity |
Sensor |
High Polymer
capacitance type |
|
Measuring range |
0 to 100%RH
(no dew condensation) |
|
Accuracy rating |
±2%RH (at 0
to 90%RH and 25°C), ±3%RH (at 90 to 95%RH and
25°C) |
|
Relative humidity/temperature
coefficient |
±0.1%RH/°C
(at the range of 5 to 80°C) |
|
Response time |
Within 15
seconds (at 90% response, 25°C and under 0.1m/s
wind flow) |
Within 1
minute (at the same condition shown
left) |
|
Output signal |
4 to 20mADC,
2-wire transmission (against 0 to
100%RH) |
|
Load resistance |
500 or lower (at 24VDC power
supply) |
|
Temperature |
Sensor |
Platinum
resistance thermometer Pt100 JIS class
A |
|
Measuring range |
0 to
50°C |
-40 to
80°C |
-40 to
180°C |
0 to
50°C |
0 to
80°C |
Accuracy rating (for 2-wire
transmission) |
±0.3°C |
±0.5°C |
±0.3°C |
|
Output signal |
4 to 20mADC,
2-wire transmission (against measuring range) for
HN-C A Pt100, 3-wire system for HN-C B |
Load resistance (2-wire
transmission) |
500 or lower (at 24VDC power
supply) |
|
Sensor interchangeability |
Plug-in
interchangeable |
Plug-in
interchangeable (including cable) |
Plug-in
interchangeable |
|
Working temperature range |
Meter |
0 to
50°C |
|
|
|
|
|
Sensor |
0 to
50°C |
-40 to
80°C |
-40 to
180°C |
0 to
50°C |
0 to
80°C |
|
Power Voltage |
12 to 24VDC
± 1V |
|
|
Connection |
Crump type
no-screw terminal connection |
Terminal
connection (M4 x 5) |
|
Material |
Meter |
Resin |
Die-cast
aluminum |
|
Protective tube |
---- |
Resin |
SUS |
|
Mounting |
Screw fixing
(M4 x 2) or DIN rail mounting |
Screw (M4
x 2) |
Flange
(JIS 5K 25A) |
|
Weight |
About
160g |
About
220g |
About
240g |
About
1.6kg |
|
Accessory |
Fixtures for
wall-mounting |
---- |
---- |
|
Others |
---- |
Cable
length: 1m, 2m, 3m (to be specified) |
---- |
---- |